top of page

Heating & Cooling Circulator

-25℃~200℃

未标题-1.png

Product Introduction

HSR series is a circulation device that can provide heat source or cold source. It has a widely application range and can be used in pharmaceutical, chemical, biological and other industries to provide heat and cold sources for reactors, liquid storage tanks, etc. It adopts fully enclosed pipeline design and high-efficiency plate heat exchanger, which reduces the demand for heat transfer fluid,  improves the heat utilization rate of the system, and realizes rapid temperature rise and fall. 

Main Features 

· Pre-cooling device, refrigeration system and heating system can be used in combination.

· It can be used with 1L~100L and other reaction kettles to maintain the materials in the reaction kettle from -25℃ to 200℃.

· Refrigeration system compressors, oil separators, solenoid valves, expansion valves, etc. are all imported brands.

· The cooling rate is fast, and rapid cooling can be achieved.

· The evaporator is a fully brazed plate heat exchanger.

· The liquid level tube can monitor oil level in real time to avoid the lack of heat transfer fluid.

· The circulatory system adopts a closed type, which is non-volatile and protects the health of the experimenter.

· The liquid storage tank and circulation pipeline are made of 304 stainless steel, which has good corrosion resistance.

· The shell adopts electrostatic plastic spray SPCC, which has good anti-corrosion effect.

· The whole machine is stable and reliable, and can run continuously for a long time.

Application Fields

Can be used in pharmaceutical, chemical, biological and other industries. Provide heat and cold source for reactors, liquid storage tanks, etc.

2.png
3.png
1.png
4.png

Customization Service

Can be used in pharmaceutical, chemical, biological and other industries. Provide heat and cold source for reactors, liquid storage tanks, etc.

5.png

High Efficiency Plate
Heat Exchanger

未标题-1.png
未标题-1.png

No oil change required.While reducing the demand for thermal fluid, improve the heat utilization rate of the system and achieve rapid temperature rise and fall. The circulation system of the heat carrier fluid is provided with an exhaust port, which can complete the filling of the heat transfer fluid

Main Specification
Model
HSR-25A
HSR-25B
HSR-25N
HSR-25T
Configuration list
operation manual, 1pc
operation manual, 1pc
operation manual, 1pc
operation manual, 1pc
Configuration list
stainless steel hose with thermal insulation (3.6 meters/piece),1pc
stainless steel hose with thermal insulation (3.6 meters/piece),1pc
stainless steel hose with thermal insulation (3.6 meters/piece),1pc
stainless steel hose with thermal insulation (3.6 meters/piece),1pc
Configuration list
matching valves,2 sets
matching valves,2 sets
matching valves,2 sets
matching valves,2 sets
Configuration list
matching valves,2 sets
matching valves,2 sets
matching valves,2 sets
matching valves,2 sets
Configuration list
main engine,1set
main engine,1set
main engine,1set
main engine,1set
Dimension
Carton size
610*510*870
610*510*870
610*510*870
840*760*1340
Dimension
Machine size
540*420*800
540*420*800
540*420*800
770*670*1180
Dimension
Gross Weight (Kg)
78
80
80
220
Dimension
Net Weight (Kg)
57
60
60
180
PLC(Optional)
Data export
Support U disk datas output, can be outputed temperature datas in TXT format
Support U disk datas output, can be outputed temperature datas in TXT format
Support U disk datas output, can be outputed temperature datas in TXT format
Support U disk datas output, can be outputed temperature datas in TXT format
PLC(Optional)
operation panel
7-inch Siemens color touch screen (10-inch optional)
7-inch Siemens color touch screen (10-inch optional)
7-inch Siemens color touch screen (10-inch optional)
7-inch Siemens color touch screen (10-inch optional)
PLC(Optional)
protocol
RS485 interface of modbus RTU protocol (optional RS232 Ethernet)
RS485 interface of modbus RTU protocol (optional RS232 Ethernet)
RS485 interface of modbus RTU protocol (optional RS232 Ethernet)
RS485 interface of modbus RTU protocol (optional RS232 Ethernet)
PLC(Optional)
control model
Material temperature and equipment outlet temperature can be switched automatically.
Material temperature and equipment outlet temperature can be switched automatically.
Material temperature and equipment outlet temperature can be switched automatically.
Material temperature and equipment outlet temperature can be switched automatically.
PLC(Optional)
Program programming
Addressable 5 programs with up to 40 steps each
Addressable 5 programs with up to 40 steps each
Addressable 5 programs with up to 40 steps each
Addressable 5 programs with up to 40 steps each
PLC(Optional)
control system
Siemens S7-200smartPLC
Siemens S7-200smartPLC
Siemens S7-200smartPLC
Siemens S7-200smartPLC
Recycle pump
The rated flow rate (L/min)
25
25
25
25
Recycle pump
The rated lift (m)
6
6
6
6
Recycle pump
Current (a)
0.5
1
1
0.5
Recycle pump
Power (w)
100
100
100
100
Recycle pump
type
Canned motor pump
Canned motor pump
Canned motor pump
Canned motor pump
Evaporator
External circulation interface size
G1/2 "inner thread turns to OD14mm pagoda
G1/2" outer thread
G1/2" outer thread
DN15
Evaporator
Liquid cell-Dimension(mm)
250*250
/
/
/
Evaporator
Liquid cell-Opening Dia (mm)
180
/
/
/
Evaporator
Liquid cell-Volume(L)
10
7
7
7
Evaporator
type
Internal thread copper tube
Plate heat exchange
Plate heat exchange
Plate heat exchange
Chiller
Number of fans
1
1
1
1
Chiller
Fan type
Outer axial rotor
Outer axial rotor
Outer axial rotor
Outer axial rotor
Chiller
type
air cooling
Sealed air cooling
Sealed air cooling
Sealed air cooling
Compressor
refrigerant
R404A
R404A
R404A
R404A、R23
Compressor
quantity
1
1
1
2
Compressor
P number
1
1
1
1.5*2
Compressor
Current (a)
3.4
1.5
/
/
Compressor
Power (Kw)
0.75
0.73
0.73
2.25
Compressor
brand name
Panasonic
Panasonic
Panasonic
Taikang
Compressor
type
fully closed
fully closed
fully closed
fully closed
Main Specification
temperature sensor
PT100
PT100
PT100
PT100
Main Specification
Safety protection
Overvoltage, delay, overcurrent, overheating, leakage, protector
Overvoltage, delay, overcurrent, overheating, leakage, protector
Overvoltage, delay, overcurrent, overheating, leakage, protector
Overvoltage, delay, overcurrent, overheating, leakage, protector
Main Specification
Refrigeration capacity (w)-﹣30℃
400
1000
250
2300
Main Specification
Refrigeration capacity (w)-0℃
1650
1600
1200
3150
Main Specification
Refrigeration capacity (w)-RT
2010
2200
1900
3650
Main Specification
Heating power (Kw)
2
2(2.5)
2(2.5)
3
Main Specification
Total power (Kw)
3
2.8
2.8
5.5
Main Specification
Phase number (p)
1
1
1
1
Main Specification
Frequency (HZ)
50
50
50/60
50
Main Specification
Voltage (v)
220
220
220
220
Main Specification
Optimum Ambient Humidity
60%
60%
60%
85%
Main Specification
Temperature (℃)
≤30℃
≤30℃
≤30℃
≤30℃
Main Specification
Design temperature range (℃)
-30~100℃;±1℃
-25~200℃;±1℃
-40~200℃;±1℃
-80~200℃;±1